Mission HB Haemoglobin Test Strips (100/Pack)
Features | Specifications |
Methodology | Reflectance Photometry |
Detection Principle | Methemoglobin |
Time to Results | < 15 seconds |
Memory | 1,000 tests with date/time and ID number |
Specimen Volume | 10 µL |
Specimen Type | Capillary and Venous whole blood |
Hb Measurement Range | 4.5-25.6 g/dL |
Hct Range | 13-75% |
Wavelengths | 525 nm |
PC Interface | Mini USB Port |
Calibration | Automatic |
Hb Within Run Prevision | ≤ 3% |
Hb Total Precision CV | ≤ 3% |
Accuracy | Hb 4.5-10 g/dL, ± 0.4 g/dL; Hb 10-25.6g/dL, ± 4% |
Operating Conditions | 15-30 °C (59-86 °F); ≤85% RH |
Meter Storage Conditions | 0-50 °C (32-122 °F); ≤90% RH |
Strip Storage Conditions | 2-30 °C (36-86 °F); ≤85% RH |
Strip Shelf Life | 2 years unopened canister; 3 months opened canister |
Power Source | 3 AAA Batteries or AC Adaptor |
Battery Life | 2,700 tests or 360 hours |
Automatic Shut Off | 8 minutes |
Line Leakage Current | 3 uA |
Meter Dimensions (L X W X H) | 127 mm × 58 mm × 25 mm (5.0″ x 2.28″ x 0.09″) |
LCD Dimensions (L X W) | 39 mm × 37 mm (1.54″ x 1.46″) |